Recall of Medifab Multistand 2000 standing frame

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Timoteos Oy.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    15577
  • Event Initiated Date
    2013-10-25
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Medifab Medical Fabrication, 32 Detroit Drive, Izone Business Hub, Rolleston 7675
  • Reason
    The issues is that the size 3 frame was not tested with a particular electric actuator, which, although sufficient on the smaller size one & two frames, has since been found to have failed under use with the larger frame., potential upgrade with a stronger actuator to prevent a fault occuring on the multistand 2000.
  • Action
    Product to be modified