Recall of Medical Intelligence HexaPOD evo RT System

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Medical Intelligence Medizintechnik GmbH.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    22989
  • Event Initiated Date
    2018-05-17
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Elekta Pty Limited - New Zealand Branch, Level 2, Windsor Court , 128-136 Parnell Road, Parnell, AUCKLAND 1052
  • Reason
    It has been noted that upon completion of a 6d workflow, the hexapod evo module may be in a tilted state, if it is moved to the position used for 3d (non-iguide) treatments. in a subsequent 3d workflow, a possible incorrect position of the patient, due to a tilt of the hexapod, may be difficult to detect without verification (e.G. imaging). this description is not relevant for an iguide workflow (6d). the problem occurs only when you switch from a 6d to a 3d workflow without iguide.
  • Action
    Manufacturer to issue advice regarding use