Recall of LIFEPAK 20 and LIFEPAK 15 Manual Defibrillator

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Physio-Control Corporation.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    21164
  • Event Initiated Date
    2017-03-03
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Medxus Ltd, Unit 4, 100 Fitgerald Avenue, Christchurch 8011
  • Reason
    During the investigation the manufacturer of the etco2 module found that compressed air may have contaminated filters within in the etco2 module. as a result of the contamination of the filter, the etco2 display on the lifepak 15 will provide a reading of xxx. there may be solid or dashed lines present on the device display in the place of an etco2 waveform and the service led will be illuminated. in this situation the etco2 monitoring functionality will no longer be available.
  • Action
    Product to be exchanged

Device

  • Model / Serial
    Model: , Affected: LIFEPAK 15 Monitor/Defibrillator with End-Tidal CO2 option.
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    NZMMDSA