Recall of Intuitive IS3000 HD Lamp Modules

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Intuitive Surgical Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    15647
  • Event Initiated Date
    2013-11-14
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Device Technologies New Zealand, 47 Arrenway Drive, Albany, AUCKLAND 0632
  • Reason
    Manufacturer has identified that the metalized coating on the is3000 hd lamp module may be incompatible with the illuminator control board.In rare circumstances the illuminator may be damaged or the lamp module might not illuminate. there have been no reports of lamp failure associated with this issue.
  • Action
    Product to be exchanged

Device

Manufacturer