Recall of Integra Ojemann Stimulator Unit

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Integra Burlington MA Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    14663
  • Event Initiated Date
    2013-05-17
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Integra Neurosciences Pty Ltd (NZ), 5B Suite B, 331 Rosedale Road, Albany, Auckland 0632
  • Reason
    Areas of the current design of the ojemann cortical stimulator (ocs2) were found not to be compliant to iec 60601-1, contraindicating the claim in the ifu. through internal evaluation, the manufacturer found that if the optional headphone jack is in use during a surgical procedure, and a non-intended voltage (such as static electricity) comes in contact with the outer case of the unit, that voltage could be transmitted to the patient and be a potential source of injury.
  • Action
    Product to be modified