Recall of Hamilton G5 and S1 Ventilators

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Hamilton Medical AG.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    18172
  • Event Initiated Date
    2015-03-04
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: InterMed Medical Ltd, 71 Apollo Drive, Albany, AUCKLAND 1311
  • Reason
    The manufacturer has received reports of the ventilator display feezing. in these cases, the red alarm lamp on top of the ventilator monitor blinks and a high-priority audible alarm is triggered, but no alarm message is displayed. the patient's ventilation continues but the screen is no longer updated and the user can no longer operate the device.
  • Action
    Software to be upgraded

Device

  • Model / Serial
    Model: Product Code: HL-159001, Affected: , Software version: between V2.40 and V2.41
  • Manufacturer

Manufacturer