Recall of Hamilton/47244/Ventilators

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Hamilton Medical AG.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    19875
  • Event Initiated Date
    2016-03-18
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: InterMed Medical Ltd, 71 Apollo Drive, Albany, AUCKLAND 1311
  • Reason
    The analysis of the customer complaint has identied an issue related to suctioning with the use of the hamilton-g5 and hamilton-s1 ventilators. after preforming the suctioning manuoeuvre, including disconnecting the patient suctioning and then reconnecting the patient the preset pattern of the ventilation may not continue as expected. this software flaw may lead to either hyperventilation or hypoventilation.
  • Action
    Software to be upgraded