Recall of Gambro Ultrafilter U9000

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Gambro Dialysatoren GmbH & Co KG.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    17113
  • Event Initiated Date
    2014-08-05
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Gambro Pty Ltd (NZ), 5B Target Court, Glenfield, Auckland 0753
  • Reason
    The manufacturer has become aware of the potential for external fluid leakage from the ultrafilter u9000 as a result of multiple disinfections using a sodium carbonate-based solution. leakage could lead to excessive fluid removal in patients.
  • Action
    Manufacturer to issue advice regarding use

Device

Manufacturer