Recall of Fujifilm FCR Go2 Mobile X-Ray System

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Fujifilm Corporation.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    14161
  • Event Initiated Date
    2013-02-01
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Fujifilm NZ Limited, 2c William Pickering Drive, Albany, Auckland
  • Reason
    It has been confirmed that in rare cases the exposure conditions to the values initially entered from console are reset when turning on the collimator lamp while pressing the collimator hand switch (releasing the electromagnetic brake) where the exposure conditions (kv and mas values) entered from console are changed manually on the control panel of the unit.
  • Action
    Software to be upgraded

Device

  • Model / Serial
    Model: Model: FCR-MB 201, Affected: APL software versions below V1.21
  • Manufacturer