Recall of Etest COLISTIN CO 256 WW B30 & B100

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Biomerieux SA.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    20579
  • Event Initiated Date
    2016-08-30
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Medi'Ray New Zealand Ltd, 53-55 Paul Matthews Road, Albany, Auckland 0632
  • Reason
    The manufacturer was alerted about a potential underestimation of the mic values for enterobacteriaceae and acinetobacter, a complaint investigation to confirm product performance and determine. while the internal and external investigations are still ongoing, to date and based on data available, the following have been identified., for the enterobacteriaceae strains, the first results of our investigation state that the performance of etestr colistin co256 conforms if used in conjunction with the mhe aga., there is a potential of false susceptible results on etestr colistin co256 ww with some mueller hinton agar.
  • Action
    Manufacturer to issue advice regarding use

Device

  • Model / Serial
    Model: , Affected: Multiple lots