Recall of Eschman RX Powered Operating Tables

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Eschmann Holdings Ltd t/a Eschmann Equipment.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    16353
  • Event Initiated Date
    2014-03-13
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Ebos Group Ltd, 14-18 Lovell Court, Rosedale, Auckland 0632
  • Reason
    Following a reported incident it has been identified that handsets are being disassembled by users, presumably to facilitate a repair., the disassembly of the handset for any reason can compromise the seal that protects the internal circuits from fluid ingress. should fluid enter the handset; this can cause an unrequested movement of the operating table top.
  • Action
    Manufacturer to issue advice regarding use

Device

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    NZMMDSA