Recall of Elekta Integrity 3.0

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Elekta Limited (England).

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    14569
  • Event Initiated Date
    2013-04-29
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Elekta Pty Limited - New Zealand Branch, Level 2, Windsor Court , 128-136 Parnell Road, Parnell, AUCKLAND 1052
  • Reason
    During beam delivery, the tolerance values are used to do a check for unwanted movements. these values are known as "static tolerances" which are related to all the digital accelerator geometric movement axes except table column. the problem is that the "static tolerances" from the calibration files (loaded database) are used for beam delivery instead of the machine calculated values. a fault occurs if the loaded database is incorrect.
  • Action
    Software to be upgraded

Device