Recall of Elekta APEX Collimator

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Elekta Limited (England).

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    14136
  • Event Initiated Date
    2013-02-05
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Elekta Pty Limited - New Zealand Branch, Level 2, Windsor Court , 128-136 Parnell Road, Parnell, AUCKLAND 1052
  • Reason
    If apex is not installed correctly and the touchguard button is used to override the inhibit and rotate the gantry, the apex™ collimator can disconnect from the radiation head if it is not correctly attached.
  • Action
    Instructions for use to be updated

Device

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    NZMMDSA