Recall of DxH 500

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Beckman Coulter Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    21476
  • Event Initiated Date
    2017-05-31
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Beckman Coulter NZ Ltd, Suite 2, Level 4, 30 St Benedicts Street, Newton, AUCKLAND
  • Reason
    The manufacturer has become aware that when the instrument temperature (<18°c to >36.5°c) is exceeded, the dxh 500 allows sample processing and result reporting despite the instructions for use stating that the run sample is inaccessible. results generated in this scenario may be erroneous. additionally:, " the message on the results screen/printout reads instrument temperature instead of instrument temperature out of range., " the results are transmitted to the laboratory information system (lis) without the accompanying instrument temperature out of range message. using the auto-transmit option could lead to the release of erroneous results to the lis.
  • Action
    Software to be upgraded