Recall of Draeger Latex Lungs (for V-series ventilators)

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Draegerwerk AG & Co KGaA.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    16398
  • Event Initiated Date
    2014-03-18
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Draeger New Zealand Ltd, Unit 4, 24 Bishop Dunn Place, East Tamaki, Auckland 2013
  • Reason
    Manufacturer has become aware of an incident where a user suffered an acute allergic reaction after using a drager latex test lung. during investigation of the incident noted that the breathing bag that forms part of the test lung and or other latex natural rubber breathing bags are labelled "nr" (natual rubber). although this is the standard abbreviation for natural rubber this labelling does not comply with the application of international standards for medical devices.Manufacturer points out that the blue reusable breathng bags which bear the marking "nr" do contain natural rubber (latex) drager breathing bags made f latex (blue) are clearly dstngiishablefrom the latex-free version made form silicon (transparent) by their colour.
  • Action
    Instructions for use to be updated

Device

Manufacturer