Recall of Draeger Evita V500 Ventilator, with optional PS500 installed

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Draegerwerk AG & Co KGaA.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    18826
  • Event Initiated Date
    2015-07-08
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Draeger New Zealand Ltd, Unit 4, 24 Bishop Dunn Place, East Tamaki, Auckland 2013
  • Reason
    Draeger has become aware of an event in which the battery capacity of the optional ps500 of an iacs workstation critical care (evita infinity v500, evita v300 and babylog vn500 ventilators) did not last as long as expected. the batteries installed in the ps500 depleted much earlier than expected although the battery indicator showed a sufficiently charged battery. the sequential alarms, "battery low" followed by" battery depleted" were given, however, the remaining time until the ventilator shut down was not foreseeable. when the battery depleted totally, the power fail alarm was generated.
  • Action
    Product to be modified