Recall of Datascope System 98/98XT, CS100/CS100i and CS300 Intra-Aortic Balloon Pumps (IABP)

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Datascope Corporation.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    16489
  • Event Initiated Date
    2014-04-03
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Maquet Australia Pty Ltd (NZ), 12G Andromeda Cres East Tamaki, Auckland 2013
  • Reason
    The manufacturer has identified a potential mechanical failure of the fan assembly associated with the power supply. a fan assembly failure could result in the power supply overheating and cause the iabp to shut down without warning.
  • Action
    Product to be modified