Recall of Codman EDS3 Lumbar Drainage Catheter Kit II

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Codman & Shurtleff Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    17380
  • Event Initiated Date
    2014-09-19
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Johnson & Johnson (NZ) Limited, 13a Gabador Place, Mt Wellington, AUCKLAND
  • Reason
    The tubing within the system that drains csf may leak or disconnect from the joints. leakage and tube separations may result in over- or under-drainage of csf from the ventricular system or introduction of air into the ventricular system (pneumocephalus). this may result in collapsed ventricles, subdural bleeding, or an inability to properly control elevated intracranial pressure. the tubing disconnection or leakage may also increase the risk of ventriculitis. if undetected or untreated each of these events may cause severe brain injury, which may lead to coma, stroke or death.
  • Action
    Product to be returned to supplier

Device

  • Model / Serial
    Model: Model: 82-1738, Affected: All lots with expiry on or before Aug-2017