Recall of Catheter, intravascular, general-purpose

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Maquet Cardiopulmonary AG.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    22941
  • Event Initiated Date
    2018-05-07
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Getinge Australia Pty Ltd (NZ), 12 G Andromeda Crescent, East Tamaki, Auckland 2013
  • Reason
    The vent plug of the arterial cannula consists of sintered polyethylene, therefore the structure of the vent plug is porous and gas-permeable. after the de-airing process is completed the customer has to separate the vent plug from the cannula to connect the extracorporeal circuit., in some cases it has come to the manufacturers attention that the vent plug has broken during the separation from the arterial cannula.
  • Action
    Product to be returned to supplier

Device

  • Model / Serial
    "="" v-b-tooltip=""> Model: , Affected: products containing the 3/8" vent plug 70000.8256
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    NZMMDSA