Recall of Carestream DRX Evolution with DIRECTVIEW Software

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Carestream Health Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    19019
  • Event Initiated Date
    2015-08-26
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Carestream Health New Zealand Ltd, c/- Quigg Partners, Level 7, 28 Brandon St, Wellington
  • Reason
    Carestream health has identified a software issue that impacts image alignment when using the automatic stitching option for long length images. if there is no use of external fiducial markers or measurement tools (eg. radio opaque ruler) when building a composite image and the technologist relies solely on the auto stitching software, there is a risk of undetected stitching errors.
  • Action
    Software to be upgraded

Device

  • Model / Serial
    Model: Carestream DRX Evolution with DIRECTVIEW Software, Affected: Versions 5.6 & greater, and V5.7 with DR LLI Auto Stitich
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    NZMMDSA