Recall of Brilliance iCT Ingenuity CT

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Philips Medical Systems (USA).

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    20820
  • Event Initiated Date
    2016-11-09
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
  • Reason
    The patient support table top may fail to move in the horizontal direction when it is commanded to do so during ct scans, due to a potential mechanical linkage failure. the system cannot automatically detect this failure to abort the scan or to alert the user.
  • Action
    Software to be upgraded

Device

  • Model / Serial
    Model: , Affected: Brilliance iCT / Brilliance iCT SP, Brilliance Big Bore Oncology / Brilliance Big Bore Radiology, Brilliance 6 / Brilliance 10 / Brilliance16/ Brilliance 16P / Brilliance 40 / Brilliance 64,Ingenuity CT / Ingenuity Core / Ingenuity Core128
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    NZMMDSA