Recall of Biomet 3i Bellatek Encode Healing Abutment

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Implant Innovations Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    19551
  • Event Initiated Date
    2015-12-11
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Biomet 3i Australia Pty Ltd (NZ), 16 Huapai Street, Onehunga, Auckland 1061
  • Reason
    Through investigation, biomet 3i determined that a limited number of product pouches may not have been sealed during packaging. the occurrence rate to date of a pouch not being sealed during packaging is approximately 0.008%. in the event the product pouch is unsealed, the sterility of the product contained within the pouch cannot be assured and/or the component may be missing.
  • Action
    Product to be returned to supplier

Device

  • Model / Serial
    Model: EHA343, EHA344, EHA346, EHA348, EHA353, EHA354, EHA356, EHA358, EHA443, EHA444, EHA446, EHA448, EHA453, EHA454, EHA456, EHA458, EHA463, EHA464, EHA466, EHA468, EHA473, EHA474, EHA476, EHA478, EHA553, EHA554, EHA556, EHA558, EHA563, EHA564, EHA566, EHA568,, Affected: Expiry dates between Nov-2015 and Nov-2020
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source
    NZMMDSA