Recall of Biomerieux Vitek 2 Neisseria-Haemophilus Identification Card

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by BioMerieux Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    18073
  • Event Initiated Date
    2015-02-19
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Medi'Ray New Zealand Ltd, 53-55 Paul Matthews Road, Albany, Auckland 0632
  • Reason
    A discrepant inoculum density range has been identified for the neisseria haemophilus (nh) identification (id) test card in the vitek® 2 7.01 product information document (ref: 514740-1en1). in section 6 the documented densichek™ plus inoculum density range is 1.80 - 2.20 (this is incorrect). the nh product information culture requirements table (table 47) indicates the densichek™ plus inoculum density range to be 2.70 - 3.30 (this is correct).
  • Action
    Instructions for use to be updated

Device

  • Model / Serial
    Model: Catalogue Number: 416381, Affected: , Software version: 7.01
  • Manufacturer

Manufacturer