Recall of Biocartis NVIdylla™ KRAS Mutation Test

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Biocartis NV.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    18964
  • Event Initiated Date
    2015-08-11
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Abacus ALS Limited, 55b Richard Pearse Drive, Airport Oaks, Auckland 2022
  • Reason
    Two false negative results were obtained during initial laboratory validation testing in a belgian clinical laboratory. a "no mutation detected in kras codon 12,13,59,61,117,146'" result was reported on a reference material of horizon, containing 5 % allelic frequency of g12a, using the idylla™ kras mutation test, lot 2207.
  • Action
    Product to be returned to supplier