Recall of Bio-Merieux API ZYM B reagent

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by BioMerieux Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    19666
  • Event Initiated Date
    2016-01-18
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Medi'Ray New Zealand Ltd, 53-55 Paul Matthews Road, Albany, Auckland 0632
  • Reason
    A customer reported a false negative result with streptococcus group d strain using reagent api zym b. customer retested the s,pyogenes strain on diferent lots and no colour changes were detected., manufacturer's investigaton confirmed a loss of performance due to light exposure after 3 months of shelf life for api zym b which lead to false negative results on some api strups.
  • Action
    Manufacturer to issue advice regarding use

Device

  • Model / Serial
    Model: 70493, Affected: Lot Nos: 100375610, 1003772410, 1003772470, 100387970, 1003790030, 1004077210, 1004097610, 1004355410 and 1004355730
  • Manufacturer

Manufacturer