Recall of Beckman Coulter UniCel DxC Synchrom Systems No Foam Reagent

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Beckman Coulter Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    20362
  • Event Initiated Date
    2016-07-06
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Beckman Coulter NZ Ltd, Suite 2, Level 4, 30 St Benedicts Street, Newton, AUCKLAND
  • Reason
    An iincompatibility between no foam dxc reagent (ref b64130) and the dxc hydropneumatic parts associated with no foam delivery has been identified., dxc no foam reagent is intended to prevent formation of foam and/or bubbles in the dxc waste system of the hydropneumatic subsystems. over time, this incompatibility may lead to:, - breaks in no foam cap assembly, - cracks in no foam waste collector; and, - seepage outside no foam tubing, this issue may lead to cracks in the waste collector which may stop instrument operation due to vacuum error, and potentially lead to biohazardous waste exposure.
  • Action
    Manufacturer to issue advice regarding use