Recall of Beckman Coulter Synchron LXi 725 Clinical Systems

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Beckman Coulter Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    13822
  • Event Initiated Date
    2012-11-20
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Beckman Coulter NZ Ltd, Suite 2, Level 4, 30 St Benedicts Street, Newton, AUCKLAND
  • Reason
    Manufacturer has identified a potential problem in two pulley drive components in the mixer pulley assemblies of the above - listed systems. in some cases pulley drive components have broken parts or have become loose, mispositioned, or disconnected from their drive shafts. when these defects occur, the pulley may not turn at all, or may turn at an incorrect speed, resulting in inadequate washing or mixing of the contents of the reaction vessel.
  • Action
    Product to be modified

Device

Manufacturer