Recall of BC Synchron Systems Enzymatic Creatininie CR-E

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Beckman Coulter Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    19282
  • Event Initiated Date
    2015-10-21
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Beckman Coulter NZ Ltd, Suite 2, Level 4, 30 St Benedicts Street, Newton, AUCKLAND
  • Reason
    Beckman coulter has confirmed that n-acetyl cysteine (nac) used in the treatment of acetaminophen overdose may cause assay interference when administered in therapeutic concentrations., the anti-oxidant properties of nac can interfere with assays employing the trinder chemical pathways., the degree of nac interference has been demonstrated to vary depending on the analyte. of the trinder chemical pathway assays, only enzymatic creatinine (cr-e) and uric acid (uric) are affected.
  • Action
    Instructions for use to be updated