Recall of Baxter Minicap Extended Life PD Transfer Set

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Baxter Healthcare Corporation.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    16724
  • Event Initiated Date
    2014-05-19
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Baxter Healthcare Ltd, 33 Vestey Drive, Mt Wellington, AUCKLAND
  • Reason
    Baxter is adding a contraindication statement to address iodine allergy for baxter's peritoneal dialysis products which contain iodine (i.E. povidone iodine) or for which iodine use is recommended. there have been no reports of complaints associated with these products for allergic or adverse reactions to iodine. furthermore, this action is intended to ensure labelling consistency across the peritoneal dialysis portfolio.
  • Action
    Instructions for use to be updated

Device

Manufacturer