Recall of APS EVOTECH Integrated Endoscope Disinfection System

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Advanced Sterilization Products.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    15928
  • Event Initiated Date
    2013-12-20
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Johnson & Johnson Medical New Zealand Ltd, 507 Mt Wellington Highway, Mt Wellington, AUCKLAND 1060
  • Reason
    The manufacturer of evotech® endoscope cleaner and reprocessors (ecr) has discovered that the evotech® ecr may cancel cycles when reprocessing the pentax eg-3670urk flexible endoscope. the pentax eg-3670urk flexible endoscope has been qualified for use with the evotech® ecr, however it cannot currently be processed in the evotech® ecr due to incorrect flow rate parameter programming in the system that is specific to this endoscope. the minimum flow rate has been programmed at levels higher than those required for effective cleaning of this scope which leads to cycle cancellation for error of low flow.
  • Action
    Software to be upgraded

Device

Manufacturer