Recall of Analgesic Gas Blender

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by O-Two Medical Technologies Inc.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    22943
  • Event Initiated Date
    2018-05-08
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: BOC Limited, 988 Great South Rd, Penrose, AUCKLAND
  • Reason
    A recent incident with the equinox advantage led to the oxygen supply to the patient being cut off. the manufacturer has determined that the fault was caused by a disengaged piston seat in the equalizer that has now been redesigned, to ensure this problem does not reoccur.
  • Action
    Product to be modified

Device

Manufacturer