Recall of Agfa Computed Radiography NX Workstation

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Agfa-Gevaert NV.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    21092
  • Event Initiated Date
    2017-02-14
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Cass Medical Limited, 803 Paparata Road, Bombay, SOUTH AUCKLAND
  • Reason
    Due to a memory issue a warning message about unexpected error and possible incorrect data integrity is shown to the user. a reboot of the workstation computer after having received the warning message prevents the issue from occurring. in one instance the warning instructions were not followed and resulted in the image of one patient being archived in another patient's record.
  • Action
    Software to be upgraded

Device

  • Model / Serial
    Model: , Affected: Software 3.0.8950
  • Manufacturer

Manufacturer