Recall of ACS900 800 Flowtron TriPuls

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by ArjoHuntleigh AB.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    21587
  • Event Initiated Date
    2017-06-27
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Arjo New Zealand Limited, 34 Vestey Drive, Mt Wellington, AUCKLAND 1060
  • Reason
    It has been identified that in some instances the pump will constantly deliver pressure to the garment without triggering the alarm. the manufacturer has investigated the symptom encountered and has concluded that the consistent flow of pressure to the garment without deflation may lead to a potential risk of injury to patients.
  • Action
    Software to be upgraded

Device

  • Model / Serial
    Model: , Affected: Software V1.099, Serial Numbers range from 1400028031 to 1600048470
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    NZMMDSA