Recall of Abbott Architect Anti-HBc II Reagent Kit

According to New Zealand Medicines and Medical Devices Safety Authority, this recall involved a device in New Zealand that was produced by Abbott GmbH & Co. KG.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    18173
  • Event Initiated Date
    2015-02-27
  • Event Country
  • Event Source
    NZMMDSA
  • Event Source URL
  • Notes / Alerts
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Extra notes in the data
    Recalling Organisation: Abbott Diagnostics Division, Ground Floor, Building D, 4 Pacific Rise, Mt Wellington, AUCKLAND 1640
  • Reason
    The architect anti-hbc ii assay shows lower calibrator, control and sample rlu values, potentially leading to reduced specificity and increased false reactive results., internal testing showed specificity dropped to 99.2% which is below the package insert claim of 99.5% on a blood donor population. cosnequently there is the potential for false reactive patient results. note that assay sensitivity is not impacted.
  • Action
    Product to be destroyed

Device

  • Model / Serial
    Model: List Numbers: 8L44-25, 8L44-30, 8L44-35, Affected: Lot Numbers: 42072LI00, 42072LI01, 42253LI00, 42253LI01, 42255LI00, 42255LI01, 42256LI00
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • Source
    NZMMDSA