Field Safety Notices about Infusion set

According to Agency for Medical Products and Medical Devices of Croatia, this field safety notices involved a device in Croatia that was produced by Shandong Qiaopai Group Co,. Ltd.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Field Safety Notice
  • Event ID
    2013_information_about_the_preventive_recall_of_the_lot_2013_01_of_the_infusion_set_medical_device_by_the_wholesaler_ma_co_plast_doo_886
  • Date
    2013-08-01
  • Event Country
  • Event Source
    AMPMDC
  • Event Source URL
  • Notes / Alerts
    Croatian data is current through September 2015. All of the data comes from the Agency for Medical Products and Medical Devices of Croatia, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Croatia.
  • Extra notes in the data
    The Agency for Medicinal Products and Medical Devices (HALMED) has received the information from the medical device wholesaler Ma-Co Plast d.o.o, about the preventive recall of the LOT 2013-01 of the Infusion Set due to its non-conformity with the Regulation (EC) 1907/2006.
    The decision about the preventive recall has been taken in collaboration with HALMED, after the test results of the aforementioned medical device lot revealed that the content of bis(2-ethylhexyl)ftalate was above the maximum allowed level set out in the Regulation (EC) 1907/2006.
    The end users are informed in a letter to quit preventively using transfusion sets from the same manufacturer Shandong Qiaopai Group Co,. Ltd and separate them from other products before the test results have been obtained.
    Here you may read the letter to the end users.
    No adverse incident related to this medical device has been reported in the Republic of Croatia by now.

Device

Manufacturer