Safety Alert for trilogy ventilators

According to Department of Health, this safety alert involved a device in Hong Kong that was produced by Philips Healthcare.

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Date
    2014-02-19
  • Event Date Posted
    2014-02-19
  • Event Country
  • Event Source
    DH
  • Event Source URL
  • Notes / Alerts
    Hong Kong data is current through September 2018. All of the data comes from the Department of Health (Hong Kong), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Hong Kong.
  • Extra notes in the data
    Medical Device Safety Alert
  • Reason
    Medical device safety alert: philips trilogy ventilators medical device manufacturer, philips healthcare, has issued a medical device safety alert concerning trilogy ventilators with model numbers t100, t200 and t202, shipped between 31 december 2013 and 30 january 2014. the affected ventilators may contain a potentially defective component on the power management board which could affect the function of the trilogy ventilator. in the worst case scenario, failure of this component could result in loss of therapy with reduced alarm functionality. loss of therapy for a patient on ventilator for life support could result in serious injury or death. according to the manufacturer, there have been no reports of death or serious injury related to this potential problem. the manufacturer advises customers to quarantine all affected devices, retrieve affected devices in use and replace them with an unaffected trilogy ventilator. customers are advised to return all affected devices to the manufacturer who will arrange replacement of the products. according to the local supplier, the affected products were distributed in hong kong. if you are in possession of the affected products, please contact your supplier for necessary actions. posted on 19 february 2014.

Device

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    DH