Safety Alert for Profile™ Single-Use Pediatric Snare

According to Department of Health, this safety alert involved a device in Hong Kong that was produced by Boston Scientific Ltd.

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Date
    2012-05-21
  • Event Date Posted
    2012-05-21
  • Event Country
  • Event Source
    DH
  • Event Source URL
  • Notes / Alerts
    Hong Kong data is current through September 2018. All of the data comes from the Department of Health (Hong Kong), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Hong Kong.
  • Extra notes in the data
    Medical Device Safety Alert
  • Reason
    Medical device safety alert: boston scientific profile™ single-use pediatric snare medical device manufacturer, boston scientific ltd, has issued an urgent field safety notice concerning the limited number of batches of the profile™ single-use pediatric snare. the manufacturer has become aware through customer complaints that it may be difficult to fully extend the snare loop from the catheter. as a result, a minor delay to the procedure may occur. according to the manufacturer, no injury is reasonably expected to occur as a result of this issue. according to the local supplier, the affected batches were not distributed in hong kong. for details, please refer to the attached field safety notice. if you are in possession of the product, please contact your supplier for necessary actions. posted on 21 may 2012.

Device

  • Model / Serial
  • Product Description
    Medical Device Safety Alert: Boston Scientific Profile™ Single-Use Pediatric Snare
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “We take a patient-first approach to assessing the applicability of every recall and communicate to regulatory bodies in all geographies where the recalled device is sold,” Boston Scientific said in a statement to ICIJ. “We have coordinated several recalls across many countries in a timely manner,” the company said, adding that it complies with all national laws, which can often vary and require different processes for reporting information or taking action on recalls. The company said it uses a rigorous and uniform process to take action on recalls and that “when we initiate a field action (e.g. recall, safety alert), every customer who has received an affected product receives a communication that includes a letter for the physician.”
  • Source
    DH