Safety Alert for Paradym DR 8550, CRT 8750 and CRT sonR 8770

According to Department of Health, this safety alert involved a device in Hong Kong that was produced by Sorin in Italy.

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Date
    2010-06-09
  • Event Country
  • Event Source
    DH
  • Event Source URL
  • Notes / Alerts
    Hong Kong data is current through September 2018. All of the data comes from the Department of Health (Hong Kong), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Hong Kong.
  • Extra notes in the data
    Press release
  • Reason
    Software defect may cause malfunction in implantable cardioverter-defibrillator the department of health (dh) today (june 9) received notification from a local supplier, innotronik hong kong limited (innotronik), about a software problem in three models of implantable cardioverter-defibrillators (icds) manufactured by sorin in italy. the affected 3 models are paradym dr 8550, crt 8750 and crt sonr 8770. the problem is due to a rare software anomaly when dual-chamber backup pacing (i.E.Pacing to both heart atrium and ventricle) is applied. according to the supplier, approximately 2,800 units have been implanted worldwide. as of 7 june this year, sorin has received one confirmed report of defect. no patient injury or death has occurred as a result of the identified software anomaly. according to innotronik, a software update will eliminate this risk. the new software will be updated to the patient's icd in the next scheduled check-up. innotronik informed that three paradym dr 8550 icds have been implanted as single chamber backup pacing in hong kong. the risk of these patients being affected by this software anomaly is low. innotronik has set up a hotline (tel no. 2877 6787) for enquiries. a dh spokesman reminded patients with implanted icds that they should seek medical attention immediately if they felt unwell. meanwhile, the dh will inform public and private hospitals, medical associations and healthcare professionals and will also follow up with innotronik on further developments, if any. ends/wednesday, june 9, 2010.

Device

  • Model / Serial
  • Product Description
    Press release: Software defect may cause malfunction in implantable cardioverter-defibrillator
  • Manufacturer

Manufacturer