Safety Alert for m/l taper with kinectiv technology femoral stems and necks

According to Department of Health, this safety alert involved a device in Hong Kong that was produced by Zimmer.

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Date
    2015-06-23
  • Event Date Posted
    2015-06-23
  • Event Country
  • Event Source
    DH
  • Event Source URL
  • Notes / Alerts
    Hong Kong data is current through September 2018. All of the data comes from the Department of Health (Hong Kong), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Hong Kong.
  • Extra notes in the data
    Medical Device Safety Alert
  • Reason
    Medical device safety alert: zimmer m/l taper with kinectiv technology femoral stems and necks the united states food and drug administration (fda) has issued medical device safety alerts concerning m/l taper with kinectiv technology femoral stems and necks, manufactured by zimmer. the manufacturer has initiated a voluntary recall of 64 lots (752 implants total) of m/l taper with kinectiv femoral stems and modular necks due to higher than allowed cytotoxicity levels found with the product. the manufacturer found a process monitoring failure that led to higher than expected amounts of manufacturing residues left on the devices. these residues can cause serious adverse health issues including allergic reactions, pain, infections, or death. use of these products may require the need for a revision surgery to replace the affected implant. according to the local supplier, the affected products are not distributed in hong kong. for details, please refer to the fda websites: http://www.Accessdata.Fda.Gov/scripts/enforcement/enforce_rpt-product-tabs.Cfm?action=select&recall_number=z-1699-2015&w=06172015&lang=eng http://www.Fda.Gov/safety/medwatch/safetyinformation/safetyalertsforhumanmedicalproducts/ucm452012.Htm http://www.Fda.Gov/medicaldevices/safety/listofrecalls/ucm451936.Htm if you are in possession of the affected products, please contact your supplier for necessary actions. posted on 23 june 2015.

Device

  • Model / Serial
  • Product Classification
  • Product Description
    Medical Device Safety Alert: Zimmer M/L Taper with Kinectiv Technology Femoral Stems and Necks
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source
    DH