Safety Alert for Capnostream 20 and Capnostream 20p Bedside Patient Monitors

According to Department of Health, this safety alert involved a device in Hong Kong that was produced by Medtronic.

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Date
    2018-08-06
  • Event Date Posted
    2018-08-06
  • Event Country
  • Event Source
    DH
  • Event Source URL
  • Notes / Alerts
    Hong Kong data is current through September 2018. All of the data comes from the Department of Health (Hong Kong), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Hong Kong.
  • Extra notes in the data
    Medical Device Safety Alert
  • Reason
    Medical device safety alert: medtronic capnostream 20 and capnostream 20p bedside patient monitors medical device manufacturer, medtronic, has issued a medical device safety alert concerning its capnostream 20 and capnostream 20p bedside patient monitors. the manufacturer has received customer reports that the date/time, nurse call and alarm settings of capnostream 20 and capnostream 20p bedside patient monitors may reset to the factory default settings when the monitor is powered off. there have been no reports of patient injury related to this issue. investigation revealed that the cause for the reset to the factory default settings is the accelerated discharge of the internal coin cell battery. user-defined institutional default settings are not lost if the monitor is not powered off, even when the internal coin cell battery is depleted. this issue does not affect the operation of the monitor’s removable li-ion battery or any other aspect of the monitor’s operation. only the reset of user-defined institutional default settings as described are affected. the manufacturer is developing a software update that will ensure user-defined institutional default settings are not lost if the monitor is powered off, except the date/time, regardless of the internal coin cell battery charge level. this software update will be available in october 2018. affected user should download the software update from the manufacturer’s website when it is available. the manufacturer will issue an update to the operator’s manual to note that the date/time setting should be verified at power on. the manufacturer is recommending continued use of capnostream devices. users should confirm that the date/time is accurately displayed. if the date/time requires reset, all user defined settings, such as alarms and nurse call, will also require reset until the software is updated. affected users should follow the instruction in the operator’s manual to set user-defined default settings. according to the local supplier, the affected products are distributed in hong kong. if you are in possession of the products, please contact your supplier for necessary actions. posted on 6 august 2018.

Device

  • Model / Serial
  • Product Description
    Medical Device Safety Alert: Medtronic Capnostream 20 and Capnostream 20p Bedside Patient Monitors
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Source
    DH