Safety Alert Or Field Safety Notices for UNISTAT L floor stand for diagnostic X-ray and superficial X-ray therapy systems

According to Medicines and Healthcare products Regulatory Agency, this safety alert or field safety notices involved a device in United Kingdom that was produced by MECALL Srl.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Safety alert / Field Safety Notice
  • Event Number
    MDA/2009/014
  • Date
    2009-02-23
  • Event Date Posted
    2014-12-17
  • Event Country
  • Event Source
    MHRA
  • Event Source URL
  • Notes / Alerts
    Data from the United Kingdom is current through April 2019. All of the data comes from the Medicines and Healthcare products Regulatory Agency, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and the United Kingdom.
  • Extra notes in the data
    The MHRA has received a report of the X-ray tube head falling onto the table after the support cable and the backup ‘safety device’ failed simultaneously. The tube head weighs 40 kg and there is a risk of death or serious patient injury if this failure occurs during examination.The X-ray tube head is supported by a metal rope that winds over a pulley system behind the tube support covers. There is a safety device fitted, which should catch the rope if it breaks.Wear on the cable may go undetected as the current service instructions for this device do not require that the whole length of the cable is inspected nor do they require that the ‘safety device’ is inspected.The manufacturer has updated the service instructions and issued a Field Safety Notice, available to download from:X-ray, ancillary equipment and accessories - Mecall Srl - UNISTAT L floor stand.
  • Reason
    (mecall srl) an x-ray tube head has fallen onto the table after the support cable and the backup ‘safety device’ failed at the same time. (mda/2009/014).
  • Action
    Identify any affected devices. Immediately remove from service. Contact the manufacturer. Ensure that the whole length of the support cable and the ‘safety device’ are inspected following the updated service instructions and are free from defect before returning into service.