Safety Alert Or Field Safety Notices for Orgentec Alegria dsDNA IgG (ORG204G) for the diagnosis of lupus

According to Medicines and Healthcare products Regulatory Agency, this safety alert or field safety notices involved a device in United Kingdom that was produced by Orgentec Diagnostika GmbH.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Safety alert / Field Safety Notice
  • Event ID
    MDA/2013/040
  • Event Number
    CON286865
  • Date
    2013-06-17
  • Event Date Posted
    2013-06-17
  • Event Country
  • Event Source
    MHRA
  • Event Source URL
  • Notes / Alerts
    Data from the United Kingdom is current through April 2019. All of the data comes from the Medicines and Healthcare products Regulatory Agency, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and the United Kingdom.
  • Extra notes in the data
  • Reason
    An increase in sensitivity may lead to false positive results in tests for the diagnosis of lupus. this issue may be related to the strip coating. orgentec has sent a recall letter.
  • Action
    Do not use and discard any remaining strips from the affected lots. Consider the need to retest positive results obtained from the affected lots. Contact Launch Diagnostics (01474 876 402) to request replacement strips for re-testing. Clinical pathologists. Biomedical scientists.

Device

  • Model / Serial
  • Product Description
    Laboratory test kit. Orgentec Alegria dsDNA IgG (ORG204G) for the diagnosis of lupus. Lot numbers: 204G30204 and 204G30205.
  • Manufacturer

Manufacturer