Safety Alert Or Field Safety Notices for Identity SR model 5172, Identity DR model 5370 and Identity XL DR model 5376

According to Medicines and Healthcare products Regulatory Agency, this safety alert or field safety notices involved a device in United Kingdom that was produced by manufacturer #17429.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Safety alert / Field Safety Notice
  • Event ID
    MDA/2006/065
  • Event Number
    CON2025297
  • Date
    2006-11-23
  • Event Country
  • Event Source
    MHRA
  • Event Source URL
  • Notes / Alerts
    Data from the United Kingdom is current through April 2019. All of the data comes from the Medicines and Healthcare products Regulatory Agency, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and the United Kingdom.
  • Extra notes in the data
  • Reason
    Potential for inaccurate display of: battery voltage expected battery longevity elective replacement indicator (eri) status at initial programmer interrogation. st jude medical has advised mhra that a programmer software anomaly can occur when the above programmers and pacemakers are used together. the manufacturer notified their uk customers of this issue in a letter dated 12 october 2006 (see appendix). the problem, which has only affected 53 out of approximately 199,100 pacemakers, results in the display of inaccurate battery voltage, expected battery longevity, and eri status. the anomaly does not affect the pacemaker's actual battery voltage, longevity or functionality. display of inaccurate data only occurs at initial programmer interrogation. correct battery voltage measurements and longevity status etc. will only be displayed when a manual measured data reading is performed. the anomaly is the result of a software interaction where a memory location within the pacemaker changes state. when this occurs, the data read on initial interrogation (from the pacemaker's memory) is not up to date. this affects the calculations for battery voltage, remaining longevity, eri status and magnet rate and therefore the displayed information is incorrect. in addition, as pacemakers approach the elective replacement time, this could result in the reporting of inappropriately high battery voltage and remaining longevity and could result in a delay in indicating eri status. to correct this problem, st jude medical has developed a software upgrade which, when installed on the programmer, will automatically correct an affected pacemaker memory location and ensure accurate battery status information is displayed. st jude medical advises that this will take place on first pacemaker interrogation following programmer software upgrade and will not extend the time needed for the appointment. st jude medical plans to release the upgraded software in december 2006 and anticipates that all potentially affected programmers in the uk will receive the software upgrade by the end of january 2007. st jude medical has already communicated this information to all uk hospitals that have received potentially affected pacemakers and programmers. neither st jude medical nor mhra have received any reports of adverse clinical events associated with this issue.
  • Action
    See detailed actions for patient management on page 3. For patients with pacemakers within two years of the longevity published in the pacemaker manual, perform a patient follow-up as soon as practically possible, giving priority to pacing dependent patients. For patients with pacemakers not within two years of published longevity, perform follow-up at the scheduled intervals. If you already routinely perform a manual measured data reading during follow-up, there is no need for any action.

Device

  • Model / Serial
  • Product Description
    St Jude Medical pacemakers: Identity SR model 5172, Identity DR model 5370 and Identity XL DR model 5376. St Jude Medical programmers: APS III model 3500, APS III model 3510 and Merlin PCS model 3650.