Recall of TROPONINE Ic

According to L’Agence nationale de sécurité du médicament et des produits de santé (ANSM), this recall involved a device in France that was produced by BAYER DIAGNOSTICS.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Date
    2000-09-19
  • Event Country
  • Event Source
    ANSM
  • Event Source URL
  • Notes / Alerts
    French data is current through early September 2018. All of the data comes from L’Agence nationale de sécurité du médicament et des produits de santé (ANSM), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and France.
  • Extra notes in the data
  • Action
    Lot 54 TROPONINE Ic reagent (references A110631 / A110632 / A116993 / A116994, used on systems ACS: 180 and ADVIA Centaur) of the company BAYER DIAGNOSTICS was withdrawn from the market, at the initiative of the manufacturer, 04.09 .2000, following the aggregation of the magnetic particles in the solid phase on some cartridges.

Device

  • Model / Serial
  • Product Description
    in-vitro_medical_device
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    LAANSM