Recall of Latex balloons XRAY

According to L’Agence nationale de sécurité du médicament et des produits de santé (ANSM), this recall involved a device in France that was produced by BALT.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Date
    2002-10-28
  • Event Country
  • Event Source
    ANSM
  • Event Source URL
  • Notes / Alerts
    French data is current through early September 2018. All of the data comes from L’Agence nationale de sécurité du médicament et des produits de santé (ANSM), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and France.
  • Extra notes in the data
  • Action
    The company BALT withdraws from the market, as a precaution, the latex balloons XRAY reference BAL1XRAY, BAL2XRAY and BAL3XRAY (all batches) following the premature aging of the latex risk of tearing during successive inflation of the balloon and release the ball of tungsten . These balloons are intended for the embolization of arteriovenous malformations and aneurysms in the neurological and peripheral territories.

Device

  • Model / Serial
  • Product Description
    medical_device
  • Manufacturer

Manufacturer

  • Source
    LAANSM