Recall of D-Dimer Plus reagent kit

According to L’Agence nationale de sécurité du médicament et des produits de santé (ANSM), this recall involved a device in France that was produced by DADE BEHRING.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Date
    2004-12-02
  • Event Country
  • Event Source
    ANSM
  • Event Source URL
  • Notes / Alerts
    French data is current through early September 2018. All of the data comes from L’Agence nationale de sécurité du médicament et des produits de santé (ANSM), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and France.
  • Extra notes in the data
  • Action
    On December 1, 2004, Dade Behring informed users of the D-Dimer Plus reagent kit - reference OQWW of a defect in stability after reconstitution. with the withdrawal of lot 544454 (per.17.10.05). a restriction of the stability period for lots 544439 to 544453 and 544455 to 544458 .. The company Dade Behring has directly informed the recipients of the incriminated lots by means of the enclosed messages:. Lot withdrawal information (02/12/2004) (10 ko Corrective action (02/12/2004) (19 kb.) This information is intended for laboratory managers, health facility directors, and health dissemination, where appropriate, to the services concerned.

Device

  • Model / Serial
  • Product Description
    in-vitro_medical_device
  • Manufacturer

Manufacturer

  • Source
    LAANSM