Recall of ARCHITECT / Prolactin Calibrators

According to L’Agence nationale de sécurité du médicament et des produits de santé (ANSM), this recall involved a device in France that was produced by ABBOTT DIAGNOSTIC.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Date
    2002-06-18
  • Event Country
  • Event Source
    ANSM
  • Event Source URL
  • Notes / Alerts
    French data is current through early September 2018. All of the data comes from L’Agence nationale de sécurité du médicament et des produits de santé (ANSM), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and France.
  • Extra notes in the data
  • Action
    ABBOTT DIAGNOSTIC withdraws batches 76916M300 (per 31.08.02) and 84229M100 (per 28.11.02) from the market for the in vitro diagnostic medical device called ARCHITECT / Prolactin Calibrators - ref. 6C27-01 and lot 76926M300 (per.25.07.02) of the ARCHITECT device / Prolactin Controls - ref. 6C27-10, following a lowering of the prolactin concentration leading to an increase in patient results and controls. As a result: no medical impact insofar as the controls identify the drift of the results; only the above-mentioned control lot may not reveal this drift, in which case the clinical impact is not considered significant. These devices allow the determination of prolactin in serum and human plasma, on the ARCHITECT automaton .

Device

  • Model / Serial
  • Product Description
    in-vitro_medical_device
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • Source
    LAANSM