Recall Or Safety Alert for Nuvo Oxygen Concentrators, Nuvo 8 and Nuvo Lite, with serial numbers between 042-10000 and 102-09335.

According to Agencia Española de Medicamentos y Productos Sanitarios (Via FOI), this recall or safety alert involved a device in Spain that was produced by Nidek Medical Products, Inc.

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Recall / Safety Alert
  • Date
    2012-07-12
  • Event Country
  • Event Source
    AEMPSVFOI
  • Notes / Alerts
    The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
  • Extra notes in the data
  • Action
    Possibility of capacitor failure in certain Oxygen Concentrators, manufactured by Nidek Medical Products, Inc, USA, which could cause a fire risk and the interruption of supplemental oxygen supply.

Manufacturer