Recall Or Safety Alert for Lifepak 1000 semi-automatic external defibrillators with software version 2.42 or earlier

According to Agencia Española de Medicamentos y Productos Sanitarios (Via FOI), this recall or safety alert involved a device in Spain that was produced by Physio-Control, Inc..

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Recall / Safety Alert
  • Date
    2015-10-16
  • Event Country
  • Event Source
    AEMPSVFOI
  • Notes / Alerts
    The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
  • Extra notes in the data
  • Action
    Update of the recommendations due to the possible unexpected disconnection of Lifepak 1000 semi-automatic external defibrillators, with software version 2.42 or earlier, manufactured by Physio-Control, Inc., USA, due to its use with batteries with low or very low charge, which will prevent the administration of the download.

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    AEMPSVFOI