Recall Or Safety Alert for Fans SLE5000, with serial number less than 53164, Fans SLE4000, with serial number less than 42040, Power supplies FP110-410 PSU, with serial number 163746 to 167558

According to Agencia Española de Medicamentos y Productos Sanitarios (Via FOI), this recall or safety alert involved a device in Spain that was produced by SLE Ltd.

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Recall / Safety Alert
  • Date
    2013-09-27
  • Event Country
  • Event Source
    AEMPSVFOI
  • Notes / Alerts
    The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
  • Extra notes in the data
  • Action
    Possibility that the capacitors of the power supply (PSU) FP110-410 of certain "SLE5000 and SLE4000 fans", manufactured by SLE Ltd, United Kingdom, present a swelling that can cause the fan to stop

Manufacturer

  • Source
    AEMPSVFOI